Performance Evaluation of Erba H7100 with Sysmex XN-1000

Main Article Content

Shanaz Khodaiji
Vidisha Mahajan
Poonam Lalla

Abstract

Introduction: Accurate hematology analyzers are essential for the diagnosis and monitoring of hematological disorders. This study evaluates the analytical performance of the Erba H7100 against the established Sysmex XN-1000 to determine agreement, precision, and suitability for routine clinical use.
Methods: A total of 243 anonymized EDTA-anticoagulated patient samples were collected and processed within two hours. Of these, 200 samples were analyzed in parallel on both analyzers for direct method comparison, while the remaining samples were used for repeatability and precision assessments. Parameters evaluated included WBC, RBC, HGB, HCT, PLT, RET%, RET#, IG#, NRBC%, and differential counts. Statistical tools included Student’s t-test, Pearson correlation, linear regression, chi-square, and Bland–Altman analysis.
Results: Strong correlations (r ≥ 0.95) were observed for WBC, RBC, HGB, HCT, PLT, and RET%. Moderate correlations were observed for lymphocyte and monocyte counts, while IG#, Bas#, and RET# showed weaker correlations and some clinically relevant differences. Most parameters showed acceptable agreement for routine clinical use, though RET# and IG# require further optimization.
Conclusion: The Erba H7100 demonstrates analytical performance comparable to Sysmex XN-1000 for most routine hematology parameters. It is suitable for clinical implementation, with potential economic advantages. Parameters like RET#, IG# will require further refinement.

Article Details

Khodaiji, S., Mahajan, V., & Lalla, P. (2026). Performance Evaluation of Erba H7100 with Sysmex XN-1000. Journal of Hematology and Clinical Research, 016–025. https://doi.org/10.29328/journal.jhcr.1001041
Research Articles

Copyright (c) 2026 Khodaiji S, et al.

Creative Commons License

This work is licensed under a Creative Commons Attribution 4.0 International License.

Lampasso J, Vasquez J. Evaluating the reliability of modern hematology analyzers. J Clin Lab Anal. 2022;36(4):e24310. Available from: https://doi.org/10.1002/jcla.24310

Briggs C, Davis B. ICSH guidelines for the evaluation of blood cell analyzers. Int J Lab Hematol. 2020;42(5):518-534. Available from: https://doi.org/10.1111/ijlh.12201

Olinder J, Ehinger D, Liljenborg E, Herwald H, Rydén C. Plasma levels of hepcidin and reticulocyte haemoglobin during septic shock. J Innate Immun. 2020;12:1-13. Available from: https://doi.org/10.1159/000508561

Lippi G, Favaloro EJ. Advances in hematology analyzers: New parameters and diagnostic opportunities. Hematol Transfus Cell Ther. 2019;41(4):381-388. Available from: https://doi.org/10.1016/j.htct.2019.05.004

Thomas C, Segal H. Analytical performance of a new hematology analyzer in comparison with Sysmex XN-1000. J Lab Med. 2020;44(1):45-52. Available from: https://doi.org/10.1515/labmed-2019-0123

Demir M, Akinci M. Comparative performance evaluation of the Sysmex XN-1000 and a new hematology analyzer. Clin Lab. 2018;64(3):503-510. Available from: https://doi.org/10.7754/Clin.Lab.2017.170926

Redondo M, Gómez-Rubio V. Comparative study of the new hematology analyzer Erba H7100 and the Sysmex XN-1000. BMC Hematol. 2021;21(1):20. Available from: https://doi.org/10.1186/s12878-021-00221-5

Banfi G, Salvagno GL. Harmonization of hematological parameters: The role of new technologies. Clin Chem Lab Med. 2019;57(8):1094-1102. Available from: https://doi.org/10.1515/cclm-2019-0167

Ahn J, Kim S. Comparative analysis of hematology analyzer performance in clinical laboratories. J Clin Pathol. 2020;73(5):295-302. Available from: https://doi.org/10.1136/jclinpath-2019-206543

Silva RF, Oliva A. Evaluation of analytical performance of hematology analyzers in clinical settings. J Blood Med. 2018;9:223-230. Available from: https://doi.org/10.2147/JBM.S179124

Almeida S, Costa P. Statistical methods for hematological analyzer comparison. Stat Med. 2019;38(8):1465-1475. Available from: https://doi.org/10.1002/sim.8053

Patel SP, Singh M. Performance assessment of hematological parameters using advanced analyzers. Int J Hematol Res. 2019;7(2):99-106. Available from: https://doi.org/10.19080/IJHR.2019.07.555710

Plebani M, Zaninotto M. The importance of analytical performance specifications in hematology. Crit Rev Clin Lab Sci. 2020;57(3):171-178. Available from: https://doi.org/10.1080/10408363.2019.1685607

Buttarello M, Plebani M. Automated blood cell counts: State of the art. Am J Clin Pathol. 2018;130(1):104-116. Available from: https://doi.org/10.1093/ajcp/aqy062

Koepke JA, Koepke JA. Reticulocytes: Advances in analytical methods. Clin Hematol. 2019;15(5):1053-1056. Available from: https://doi.org/10.1093/clinhemat/aqz112

Al-Sweedan S, Kheir E. Evaluating the performance of hematology analyzers for immature granulocyte counts. Pediatr Hematol Oncol. 2020;37(4):235-241. Available from: https://doi.org/10.1080/08880018.2020.1729273

Johnson WL, Kim S. Addressing discrepancies in hematology analyzer results: A statistical approach. Biostat Biometrics Open Access J. 2020;6(1):112-119.

Bain BJ. Blood cells: A practical guide to laboratory hematology. 5th ed. Wiley-Blackwell; 2019.

Briggs C, Culp N, Davis B. ICSH guidelines for the evaluation of blood cell analyzers, including those used for differential leucocyte and reticulocyte counting. Int J Lab Hematol. 2012;34(5):518-534. Available from: https://doi.org/10.1111/j.1751-553X.2012.01408.x

McKenzie SB, Williams JL, Landis-Piwowar K. Clinical laboratory hematology. 3rd ed. Pearson; 2019.

Buttarello M, Plebani M. Automated blood cell counts: State of the art. Am J Clin Pathol. 2018;130(1):104-116. Available from: https://doi.org/10.1093/ajcp/aqy062

Koepke JA, Koepke JA. Reticulocytes: Advances in analytical methods. Clin Hematol. 2019;15(5):1053-1056. Available from: https://doi.org/10.1093/clinhemat/aqz112

Al-Sweedan S, Kheir E. Evaluating the performance of hematology analyzers for immature granulocyte counts. Pediatr Hematol Oncol. 2020;37(4):235-241. Available from: https://doi.org/10.1080/08880018.2020.1729273

Johnson WL, Kim S. Addressing discrepancies in hematology analyzer results: A statistical approach. Biostat Biometrics Open Access J. 2020;6(1):112-119. Available from: https://doi.org/10.19080/BBOAJ.2020.06.555678

Almeida S, Costa P. Statistical methods for hematological analyzer comparison. Stat Med. 2019;38(8):1465-1475. Available from: https://doi.org/10.1002/sim.8053

Patel SP, Singh M. Performance assessment of hematological parameters using advanced analyzers. Int J Hematol Res. 2019;7(2):99-106. Available from: https://doi.org/10.19080/IJHR.2019.07.555710